New TAM Air

The isothermal calorimeter of choice for labs that require high sensitivity and the industry’s best baseline stability for both R&D applications and long term process analysis.

View brochure Contact Us

The new TAM Air from TA Instruments is a powerful isothermal calorimeter with heat flow sensitivity in the microwatt range and unmatched baseline drift performance.  New TAM Air Advantages include:

  • Operating temperature range: 5 °C – 90 °C
  • Circulating air thermostat with temperature stability at ±0.02 °C
  • Updated electronics for accurate data acquisition and EU ROHS compliance
  • User choice of 8-channel or 3-channel calorimeter configuration
  • Improved baseline drift performance for long term (e.g. weeks) experiments
  • Connect up to 8 external probes including I/O Voltage card
  • Compatible with all current TAM Air ampoules
  • Operates with complete TAM IV Lab Asst. software

All calorimetric channels are of the twin type, consisting of a sample and a reference chamber. The 8 channel calorimeter will accommodate ampoules with a 20 mL volume. The 3 channel calorimeter will accommodate ampoules with a 125 mL volume. The TAM Air is the recognized instrument of choice in the research and development, as well as quality control process of cement and concrete. With the higher volume ampoules in the 3-channel calorimeter, the TAM Air can also be utilized in food and environmental science applications

Thermostat

The TAM Air is an air-based thermostat utilizing a heat sink to conduct heat away from the sample and effectively minimize outside temperature effects.

Thermostat Specifications
Calorimeter Positions 8 or 3
Operating Temperature Range 5 ºC to 90 ºC
Thermostat type Air
Thermostat Stability ± 0.02º C over 24 hours

 

Two user changeable calorimeter blocks are available for the TAM Air thermostat. Both the 3 channel large volume and 8 channel standard volume calorimeter consist of a calorimeter block and data logging system that is required for use with the TAM Air thermostat. All calorimeters are twin-type (sample and reference).

Eight (8) Channel Standard Volume Calorimeters

Eight (8) Channel Standard Volume Calorimeters

TAM Air calorimeter

The 8 channel standard volume calorimeters are twin-type (sample and reference) and designed for use with 20 mL glass or plastic ampoules or the 20 mL Admix ampoules.

8 Channel Standard Volume Calorimeter
Short term noise <±4 µW
Baseline Drift over 24 hours < 25µW*
Limit of detection 4 µW

* Baseline drift specification is based on a 24-hour room temperature cycle and can be extended to be valid for multiple days and up to several weeks.

Three (3) Channel Large Volume Calorimeters

Three (3) Channel Large Volume Calorimeters

TAM Air calorimeter

3 channel large volume calorimeters contain three twin style calorimeters that will accommodate either 125 mL glass ampoules or 125 mL stainless steel ampoules.

3 Channel Large Volume Calorimeter
Short term noise <±8 µW
Baseline Drift over 24 hours < 55µW*
Limit of detection 8µW

* Baseline drift specification is based on a 24-hour room temperature cycle and can be extended to be valid for multiple days and up to several weeks.

 

Ampoules

A wide variety of closed ampoules are available for the TAM Air Calorimeters which enables maximum flexibility for sample management and maximum sensitivity. All 20 mL closed ampoules currently available from TA Instruments can be used in the 8 Channel TAM Air Calorimeter, as well as the 20 mL Admix ampoules. The 3 Channel Calorimeter will accommodate either 125 mL glass ampoules or 125 mL stainless steel ampoules.

Eight Channel Calorimeter

Eight (8) Channel Calorimeter

TAM Air ampoule

20 mL CLOSED AMPOULE
20 mL ampules are available in glass, stainless steel and Plastic (HDPE ).

TAM Air ampoule

20 mL ADMIX AMPOULE

The Admix Ampule is a 20mL accessory available to initiate reactions inside the calorimeter and can be configured with or without a motor for stirring.


Three Channel Calorimeter

Three (3) Channel Calorimeter

TAM Air ampoule

125 mL POLISHED GLASS AMPOULE
Polished glass ampoules are preferred for wet or moisture-containing samples. The ampoules have a polished inner surface at the top to give a proper seal to the o-ring in the metal lid. The ampoules are disposable or may be washed for reuse.

TAM Air ampoule

125 mL STAINLESS STEEL AMPOULE
Stainless steel ampoules are suitable for most samples that can be cleaned out after measurements. These ampules are sealed with a circlip and a Viton o-ring.

Experiments

The TAM Air system is especially well-suited to processes that evolve or consume heat over the course of days and weeks such as cement and concrete hydration, food spoilage, microbial activity and more.


Cement Setting Times with Admixtures

Cement Setting Times with Admixtures

TAM Air

Cement Setting Times with Admixtures When testing cement lots without admixture little variation in setting time is observed. When different lots of cement are tested with the same admixture, large differences in setting times can be observed.


Cement Thermal Profile with Contaminants

Cement Thermal Profile with Contaminants

TAM Air

Cement Thermal Profile with Contaminants A steady decrease in thermal power is evident as the contamination of the cement mortar by a mixture of soil and sawdust increases(0; 0.9; 2.5 and 5.9% of w/c=0.6 cement mortar).

Cement Sulfate Depletion

Cement Sulfate Depletion

TAM Air

Cement Sulfate Depletion The sulfate depletion peak occurs after the main silicate peak. This can be used for a rapid indication of the optimum sulfate content of a laboratory ground clinker. The results indicate that 2.5% added S03 might be sufficient to bring the resulting Portland cement to optimum S03 level.


Cement Hydration - Admix Ampoule

Cement Hydration – Admix Ampoule

microcal image

Hydration of calcium sulfate hemihydrate with Admix Ampoule. 2 g Calcium Sulfate Hemihydrate samples was hydrated with distilled water (blue) or hydrated with a 5% Sodium Chloride solution (red). When NaCl is added to the hydration solution, the time it takes for the calcium sulfate hemihydrates to complete the hydration reaction is reduced by over one-half.

 


The new TAM Air now takes advantage of all the instrument operating features, data acquisition features and data analysis features in the full version of TAM IV Lab Assistant software.   The new TAM Air is not compatible with previous version of TAM Air Assistant software.

Download TAM IV Lab Asst. Software

TAM Air Photo Gallery